Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia
The Philips Intellivue MP50 monitor provides a method for non-invasive, near-continuous blood pressure (BP) monitoring and is designed to be an alternative to direct intra-arterial BP (IABP) measurement. However, no studies have specifically compared non-invasive and invasive BP measurements using the monitor. The present retrospective study observed 515 patients undergoing surgery with general anesthesia, whose invasive (intra-radial, femoral or dorsalis pedis artery) and non-invasive (oscillometric) BP (NIBP) were monitored simultaneously using the monitor. These data were analyzed using correlations, regressions and Bland-Altman plots. The patients were placed in a supine position during surgery. The correlation data for invasive BP and NIBP measurements were: for intra-radial measurements, r(2)=0.51 (bias and precision, 11.04 +/- 15.22 and 14.76 +/- 11.64 mmHg, respectively) for systolic BP (SBP) and r(2)=0.27 (6.17 +/- 11.95 and 9.77 +/- 9.25 mmHg, respectively) for diastolic BP (DBP); for intra-femoral measurements: r(2)=0.57. (14.79 +/- 14.55 and 17.15 +/- 11.68 mmHg, respectively) for SBP and r(2)=0.45 (4.12 +/- 9.70 and 7.49 +/- 7.40 mmHg, respectively) for DBP; and for intra-dorsalis pedis measurements: r(2)=0.33 (13.00 +/- 16.81 and 17.34 +/- 12.28 mmHg, respectively) for SBP and r(2)=0.30 (0.17 +/- 11.27 and 8.44 +/- 7.46 mmHg, respectively) for DBP. According to this data, the NIBP measured by the Philips Intellivue MP50 monitor showed low positive correlations and poor agreement with the IABP, as calculated by Bland-Altman analysis. Therefore, the use of oscillometric BP measured by the monitor in surgery patients under general anesthesia is not generally recommended.
基金:
National Natural Science Foundations of China [30901404]
第一作者单位:[1]Huazhong Univ Sci & Technol,Tongji Med Coll,Tongji Hosp,Dept Urol,Wuhan 430030,Hubei,Peoples R China
通讯作者:
通讯机构:[2]Huazhong Univ Sci & Technol,Tongji Med Coll,Tongji Hosp,Dept Anesthesiol,Wuhan 430030,Hubei,Peoples R China[*1]Huazhong Univ Sci & Technol,Tongji Med Coll,Tongji Hosp,Dept Anesthesiol,1095 Jiefang Ave,Wuhan 430030,Hubei,Peoples R China
推荐引用方式(GB/T 7714):
Meng Xianghu,Zang Guanghui,Fan Longchang,et al.Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia[J].EXPERIMENTAL AND THERAPEUTIC MEDICINE.2013,6(1):9-14.doi:10.3892/etm.2013.1121.
APA:
Meng, Xianghu,Zang, Guanghui,Fan, Longchang,Zheng, Lei,Dai, Jinzhen...&Zhang, Chuanhan.(2013).Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia.EXPERIMENTAL AND THERAPEUTIC MEDICINE,6,(1)
MLA:
Meng, Xianghu,et al."Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia".EXPERIMENTAL AND THERAPEUTIC MEDICINE 6..1(2013):9-14