Analytical validation and investigation on reference intervals of aldosterone and renin in Chinese Han population by using fully automated chemiluminescence immunoassays

文献详情

资源类型:
WOS体系:
Pubmed体系:

收录情况: ◇ SCIE

单位: [1]Huazhong Univ Sci & Technol, Tongji Hosp, Tongji Med Coll, Dept Lab Med, Jiefang Ave 1095, Wuhan, Hubei, Peoples R China [2]Huazhong Univ Sci & Technol, Tongji Hosp, Tongji Med Coll, Dept Phys Examinat Ctr, Wuhan, Hubei, Peoples R China
出处:
ISSN:

关键词: Plasma direct renin concentration Aldosterone Primary aldosteronism Plasma renin activity Reference intervals Aldosterone renin ratio

摘要:
Background: The aldosterone/renin ratio (ARR) is recommended to screen for primary aldosteronism (PA) in hypertension. We estimated fully automated chemiluminescence immunoassays (CLIA) for plasma aldosterone concentrations (PAC) and plasma direct renin concentrations (PRC) and investigated their reference intervals in Chinese Han population. Methods: PAC and PRC were measured on a fully automated analyzer (LIAISON XL, DiaSorin, Italy). Performance characteristics were estimated according to CLSI approved guidelines. 328 healthy individuals were selected for reference intervals investigation. Results simultaneously tested by CLIA and radioimmunoassays were reviewed from 123 patients with hypertension and/or adrenal space-occupying lesion. PAC/PRC ratio (ARR(prc)) was compared to PAC/plasma renin activity (PRA) ratio (ARR(pra)). Results: Within-laboratory imprecision was 5.6%-6.7% for PAC and 3.0%-3.3% for PRC. The LoQ was 72.2 pmol/L for PAC and 1.27 mIU/L for PRC. Linearity was excellent in the range of concentrations between 94 and 2708 pmol/L for PAC and 1.3-461.8 mIU/L for PRC. Interferences of hemoglobin, unconjugated bilirubin and lipaemia could be acceptable, but not of conjugated-bilirubin when renin and aldosterone at low concentrations. The central 95% reference intervals for males: PAC: 76-722 pmol/L, PRC: 3.3-92.7 mIU/L, ARR: 2.2-46.0 pmol/mIU; for females: PAC: 85-1010 pmol/L, PRC: 3.7-99.8 mIU/L, ARR: 3.6-68.4 pmol/mIU. Upper reference limits for ARR of younger and older men were lower than women. ARR(prc) and ARR(pra) showed almost perfect agreement (kappa=0.815) for screening PA. Conclusion: The DiaSorin tests are valuable analytical options for PAC and PRC measurements. We recommend sex-specific and age-specific reference intervals of these items should be estimated.

语种:
被引次数:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2017]版:
大类 | 3 区 医学
小类 | 3 区 医学实验技术
最新[2025]版:
大类 | 3 区 医学
小类 | 3 区 医学实验技术
JCR分区:
出版当年[2016]版:
Q2 MEDICAL LABORATORY TECHNOLOGY
最新[2023]版:
Q2 MEDICAL LABORATORY TECHNOLOGY

影响因子: 最新[2023版] 最新五年平均 出版当年[2016版] 出版当年五年平均 出版前一年[2015版] 出版后一年[2017版]

第一作者:
第一作者单位: [1]Huazhong Univ Sci & Technol, Tongji Hosp, Tongji Med Coll, Dept Lab Med, Jiefang Ave 1095, Wuhan, Hubei, Peoples R China
通讯作者:
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:428 今日访问量:0 总访问量:412 更新日期:2025-04-01 建议使用谷歌、火狐浏览器 常见问题

版权所有:重庆聚合科技有限公司 渝ICP备12007440号-3 地址:重庆市两江新区泰山大道西段8号坤恩国际商务中心16层(401121)